EasyManua.ls Logo

Hach BioTector B3500C - Data Program; Information

Hach BioTector B3500C
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 110
8.2.10 Data Program
MMC/SD CARD
This menu allows the user to program the output device
communication port configuration profiles for MMC/SD flash
card.
PRINT MODE ENGINEERING
PRINT MODE specifies the format in which the data will be sent
over the communication port to all output devices. The available
modes are STANDARD (includes time, date and calibrated
analysis data etc.), and ENGINEERING (includes un-calibrated
data, temperature and atmospheric pressure, pump diagnostic
information, liquid phase measurement information etc. in
addition to the STANDARD data).
REACTION ON-LINE YES
REACTION ON-LINE enables (YES by default) or disables (NO)
the automatic printout of the reaction result at the end of each
reaction.
FAULT ON-LINE YES
FAULT ON-LINE enables (YES by default) or disables (NO) the
automatic printout of a fault or warning log as soon as it occurs.
CONTROL CHARS NO
CONTROL CHARS enables (YES) or disables (NO by default)
the control characters sent with the RS232 data for the output
device operation.
DECIMAL POINT
DECIMAL specifies whether a POINT (.) by default or a COMMA
(,) character represents the decimal point in the downloaded
analysis data files for all output devices.
8.2.11 Information
CONTACT INFORMATION
Contact menu displays the contact details of the manufacturer
and if programmed the distributor.
SOFTWARE
VERSION 02.01.00
VERSION displays the system firmware (software) revision
number.
DATE May 5 2011
DATE displays the firmware (software) revision date.
IDENTIFICATION
B5C 2.01 247
IDENTICIFATION displays the specific identification number
(system serial number) for the BioTector.