5.3.2 Reagent Connections
Special precautions are needed when working with chemical reagents, both
when renewing reagents and when dealing with leaks or spills. Some reagents
can cause chemical burns and may cause injury or death if swallowed. Please
refer to the symbols and codes on the reagent containers.
Use of 20 or 25 liter containers is recommended for each BioTector reagents. Figure 13 below shows the
correct setup and connection of BioTector reagents: Acid (6.0 N Sulfuric Acid containing 350 mg/l Manganese
Sulphate Monohydrate) and Base (1.2 N Sodium Hydroxide). See Section 6 Reagents and
Calibration Standards for further details.
Figure 13 BioTector reagent setup and connections