5.2 Unpacking and Installation
The BioTector analyzer weighs typically 46kg (101 lb). Therefore, appropriate
means should be used for transporting, un-packing, lifting and installing the
BioTector.
The BioTector analyzer is shipped ready to be installed, with a kit of parts including sample tubes, reagent
tubes and a selection of spare parts, spare fuses and ferrules.
When the BioTector shipping container is opened, it must be inspected against the shipping list located inside
the container. Additionally, it should be confirmed that no damage was caused to the BioTector during
shipment.
Any issues must be reported to the manufacturer within 3 days.
The BioTector is shipped with a Commissioning and Startup checklist (see Section 7 Analyzer
Commissioning and Startup for details). In order to ensure a quick and trouble free installation, this list should
be followed in the correct sequence.
Points to note regarding installation:
The BioTector should be located as close to the sample point as possible.
Unless it is specified, the BioTector has an Ingress Protection rating of IP54. It is recommended that
the BioTector is installed in a dry, well ventilated and dust free area.
The BioTector should be installed where the ambient temperature is between 5 and 45°C. If the
ambient temperature exceeds 45°C, a vortex cooler can be installed to reduce the internal
temperature of the BioTector.
The BioTector should be installed vertically, with the maximum variation on each axis less than 2°.
Ensure that there is enough clearance at the front of the BioTector to allow the door to be opened.
Ensure that there is enough clearance at the left hand side of the BioTector for the tube and electrical
connections. There should be enough clearance at both sides for the cooling fan to operate
unimpeded.
If there are corrosive gasses in the area, then the BioTector fan should be blanked off,
and an instrument air purge system should be fitted.
Caution
Caution

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B3500C and is the answer not in the manual?

Hach BioTector B3500C Specifications

General IconGeneral
BrandHach
ModelBioTector B3500C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals