SYSTEM PROGRAM 2 TC
SAMPLE IN 5p , P-V
SAMPLE IN defines the amount of sample (in pulses) to be
injected into the reactor in TC mode at range 2.
SYSTEM PROGRAM 2 VOC
All system parameters for SYSTEM PROGRAM 2 VOC analysis
mode are covered in System Program 2 TC and System
Program 2 TIC + TOC above.
SYSTEM PROGRAM 3
SYSTEM PROGRAM 3 TIC + TOC
RANGE CHANGE 2-3 150.0 mgC/l
RANGE CHANGE 2-3 defines the concentration level for the
BioTector to change its range automatically from range 2 to
range 3.
The RANGE CHANGE 2-3 is typically the top concentration
point of TOC RANGE 2 in System Range Data menu by default.
RANGE CHANGE 3-2 120.0mgC/l
RANGE CHANGE 3-2 specifies the concentration level for the
BioTector to change its range automatically from range 3 to
range 2.
The RANGE CHANGE 3-2 is typically 20% less of the top
concentration point of TOC RANGE 2 in System Range Data
menu by default.
RANGE CHANGE 3-1 40.0mgC/l
RANGE CHANGE 3-1 defines the concentration level for the
BioTector to change its range automatically from range 3 to
range 1.
The RANGE CHANGE 3-1 is typically 20% less of the top
concentration point of TOC RANGE 1 in System Range Data
menu by default.
SAMPLE IN 3p P-V
SAMPLE IN states the amount of sample (in pulses) to be
injected into the reactor at range 3.
TIC ACID 12p
TIC ACID defines the amount of acid reagent (in pulses) to be
injected into the reactor at range 3.
BASE 70p
BASE specifies the amount of base (in pulses) to be injected into
the reactor for the BASE OXIDATION phase at range 3.
BASE OXIDATION 130s
BASE OXIDATION defines the oxidation time (150s by default),
which can be programmed independently for range 3.
If the Base Oxidation time is different than the one programmed
in System Program 1 menu, depending on the difference
between the two settings, the reaction time at range 3 will be
either longer or shorter than the reaction time displayed in
System Program 1 menu.
TOC ACID 58p
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase at range 3.