20. TC Cal Std. In VOC systems, the sum of the TIC and TOC Calibration Standard solution is displayed as
TC Calibration Standard. When TOC Calibration Standard is programmed as 0.0mgC/l, and when a
concentration of TIC Calibration Standard is programmed above, BioTector displays the TC Calibration
Standard as 0.0mgC/l on purpose. This allows the calibration of TIC without any effect on the TC
calibration. See definitions for TOC Calibration Standard above, Section 8.3.4.3 Span Program and
Section 6.2 Calibration Standards for further details.
21. TC Check Std. In VOC systems, the sum of the TIC and TOC Check Standard solution is displayed as
TC Check Standard. When TOC Check Standard is programmed as 0.0mgC/l, and when a concentration
of TIC Check Standard is programmed above, BioTector displays the TC Check Standard as 0.0mgC/l on
purpose. This allows the check of TIC without any effect on the TC check.
23. Span Program. Span Program is a link to Maintenance, System Configuration, Sequence Program,
Span Program menu (see Section 8.3.4.3 Span Program).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B3500C and is the answer not in the manual?

Hach BioTector B3500C Specifications

General IconGeneral
BrandHach
ModelBioTector B3500C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals