SYSTEM PROGRAM 2
SYSTEM PROGRAM 2 TIC + TOC
RANGE CHANGE 1-2 50.0µgC/l
RANGE CHANGE 1-2 defines the concentration levels for the
BioTector to change its range automatically from range 1 to
range 2.
The RANGE CHANGE 1-2 is typically the top concentration
point of TOC RANGE 1 in System Range Data menu by default.
RANGE CHANGE 1-3 150.0µgC/l
RANGE CHANGE 1-3 specifies the concentration level for the
BioTector to change its range automatically from range 1 to
range 3.
The RANGE CHANGE 1-3 is typically 50-75% of the top
concentration point of TOC RANGE 2 in System Range Data
menu by default.
RANGE CHANGE 2-1 40.0µgC/l
RANGE CHANGE 2-1 states the concentration level for the
BioTector to change its range automatically from range 2 to
range 1.
The RANGE CHANGE 2-1 is typically 20% less of the top
concentration point of TOC RANGE 1 in System Range Data
menu by default.
SAMPLE IN 5p P-V
SAMPLE IN identifies the amount of sample (in pulses) to be
injected into the reactor at range 2.
TIC ACID 12p
TIC ACID defines the amount of acid reagent (in pulses) to be
injected into the reactor at range 2.
BASE 70p
BASE specifies the amount of base (in pulses) to be injected into
the reactor for the BASE OXIDATION phase at range 2.
BASE OXIDATION 90s
BASE OXIDATION defines the oxidation time (90s by default),
which can be programmed independently for range 2.
If the Base Oxidation time is different than the one programmed
in System Program 1 menu, depending on the difference
between the two settings, the reaction time at range 2 will be
either longer or shorter than the reaction time displayed in
System Program 1 menu.
TOC ACID 58p
TOC ACID states the amount of acid (in pulses) to be injected
into the reactor for the TOC SPARGE phase at range 2.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B3500ul and is the answer not in the manual?

Hach BioTector B3500ul Specifications

General IconGeneral
BrandHach
ModelBioTector B3500ul
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals