8.3.7 Result Integration
TOC INTEGRATION 3
TOC INTEGRATION (3 by default) defines the number of
reaction results to be averaged to get the actual TOC result.
TOC INT. LIMITS 10% , 25
TOC INTEGRATION LIMITS controls the averaging function
defined in TOC INTEGRATION above. The first parameter
“10%” defines the % band variation and the second parameter
0.025 defines the absolute variation in µgC/l. If the reaction
result is outside the specified bands (TOC INTEGRATION
LIMITS), then the reaction result averaging is omitted. In other
words, the result, which is outside the TOC INTEGRATION
LIMITS, is displayed at the end of the reaction without
averaging. If the result is within the TOC INTEGRATION LIMITS,
the averaging function is continued and the programmed number
of reaction results (defined in TOC INTEGRATION above) is
averaged.