Section 6 Reagents and Calibration Standards
6.1 Reagents
BioTector TOC analyzer uses the following reagents:
I. Acid: 1.8 N Sulfuric Acid (H
2
SO
4
) Reagent containing 80 mg/l Manganese Sulfate Monohydrate
II. Base: 1.2 N Sodium Hydroxide (NaOH) Reagent
Reagents should not contain high levels of organics, nitrates and phosphates. Ideally, the level of organics,
nitrates and phosphates should be less than 100 g/l (ppb) in the deionized water used to prepare TOC
analyzer reagents.
Acid and Base reagents are stable up to 1 year. The reagents must be stored in a safe and secure location,
where the temperature does not drop below 2°C, in compliance with the local regulations. Table 6 below
summarizes the total days each BioTector TOC Analyzer reagent lasts at various system configurations:
Table 6 BioTector TOC Analyzer Reagent Consumption
REAGENTS
Container Size
(Liters)
TOTAL DAYS REAGENT LASTS
Acid
19
122
Base
19
122
Above table is derived from several online operation parameters such as 100% online time.
Recommended bunds (reagent spill trays) to contain above quantity reagents are 1x 50 Liters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B3500ul and is the answer not in the manual?

Hach BioTector B3500ul Specifications

General IconGeneral
BrandHach
ModelBioTector B3500ul
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals