Page 36
MONITORING RANGES:
TOC
Standard Range 0-5000µgC/l
Accuracy and Repeatability 2% of reading or 10 µgC/l, whichever is greater
Limit of Detection 10 µg/l
SAMPLE & ENVIRONMENTAL CONDITIONS
Sample Volume: Up to 12.0ml
Sample Inlet Pressure: Ambient (for applications with high sample pressure, sampling systems are
available)
Drain Pressure: Ambient (for applications with high drain pressure, optional systems are
available)
Sample Inlet Temperature: 2°C – 60°C (36°F - 140°F)
Sample Flow Rate: Minimum 100ml per sample
Sample Particle Size: Up to 100 microns
Ambient Temperature: 5°C – 45°C (41°F - 113°F)
Cooling and heating options are available.
Humidity: 5% - 85%, non-condensing
Ingress Protection: IP44, standard fan cooled, maximum ambient temperature 45°C (113°F)
IP54, air cooled, maximum ambient temperature 35°C (95°F)
IP54, vortex cooled, maximum ambient temperature 50°C (122°F)
System Sound: < 60 dBa
The manufacturer has a continuous research and development program. Specifications may
therefore be changed without notice. For specification updates, please contact the
manufacturer.