Figure 17
Example Of A User Program Printout
DR/2010 User Program Report SN:789456123
Printed 26-MAR-96 11:32:55
Adj Chloride Test - Program Number: 968 Last Calibrated: 23-JAN-96 09:21:00
Default Measurement Form: 00.00 mg/L Cl
Alternate Form 2: 0.000 mg/L NaCl Factor: 0.1420
Alternate Form 3: 0.000 mg/L EDL Factor: 1.000
High Limit: 8.97 mg/L Cl
Low Limit: 0.05 mg/L Cl
Monochromator Wavelength: 420 nm
Standard Concentration Absorbance
1 0.05 mg/L Cl 0.007
2 0.20 mg/L Cl 0.028
3 0.50 mg/L Cl 0.070
4 1.00 mg/L Cl 0.140
5 2.00 mg/L Cl 0.281
6 5.00 mg/L Cl 0.700
7 10.00 mg/L Cl 1.400
Calibration Curve Forced Through Zero? Yes
Calibration Style: C = a + bA + cA
2
+dA
3
a= 0.000
b = 0.1050
c = 0.0052
d = - 0.0002
r
2
= 0.9994
Timer 1: 3:30 Timer 2: 5:00
3.8 Erasing Data
You may only erase stored data after it is recalled using the
RECALL key. To
recall data, see 3.6, Recalling Data. Erasing stored data may take up to 35
seconds. The instrument will not display an “in progress” message. Please
leave the instrument on during the process.
3.8.1 Erasing Single Data Points
1. When the desired group and sample number are displayed, press
SHIFT
ERASE
. The display will show (i.e., G100 is the group number
assigned; #3 is the sample number):
2. Press
ENTER. The display will show ERASED and then return to the
displayed group and sample number.
nm
+8.8.:8.8.
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
888540
*0.16
Erase G100 #3