79
4. To delete the program, press
ENTER (if not, press EXIT). The display will show:
5. If the number displayed is the appropriate program number, press
ENTER. If not, use the arrow keys to scroll through the other User-
entered program numbers, or enter the appropriate number using the
NUMERIC keys. Press ENTER when the appropriate number is displayed.
6. The display will show:
7. Press
ENTER to erase the program. The display will show:
andreturntothe
Delete Program Menu.
3.13 Copying a User-Entered Program
The Copy Program submenu allows users to copy user-entered methods
or templates for methods that require a user calibration. When you copy a
program or template, all the setup selections, timer information and
calibration data is copied.
Factory-stored templates have numbers between 900 and 949. The
calibration data in the templates may be nonsensical numbers and need to
be replaced with actual calibration data. For other templates, the
calibration data will yield semi-accurate results, but should be replaced
for more accurate results.
Programs that have a template for copying are arsenic, barium,
autocatalytic copper, potassium, TKN and sulfate.
Only user-entered programs and templates may be copied. Hach programs
may not be copied.