UV Lamp Check
To verify the ultraviolet lamp (normal life equals 5000 hours) is working
properly, perform the following test:
1. Prepare a 5.0 mg/L benzotriazole standard solution by pipetting 10.0 mL of
Benzotriazole Standard Solution, 500-mg/L benzotriazole, into a 1-liter
volumetric flask. Dilute to volume.
2. Analyze according to the above procedure. If the result is significantly below
5 mg/L, replace the lamp.
Method Performance
Precision
Standard: 4 mg/L benzotriazole or tolyltriazole
See Section 3.4.3 Precision on page 44 for more information, or if the standard
concentration did not fall within the specified range.
Sensitivity
See Section 3.4.5 Sensitivity on page 44 for more information.
Summary of Method
Benzotriazole or tolyltriazole, used in many applications as corrosion inhibitors
for copper and copper alloys, are determined by a proprietary catalytic
ultraviolet (UV) photolysis procedure requiring less than 10 minutes to perform.
Test results are measured at 425 nm.
Required Reagents
Quantity Required
Description per test Unit Cat. No.
Triazole Reagent Powder Pillows...........................................................1 pillow........... 100/pkg.............21412-99
Required Apparatus
UV Safety Goggles....................................................................................1 ....................... each....................21134-00
Sample Cells, 10-20-25 mL, w/cap.........................................................2....................... 6/pkg.................24019-06
Select one based on available voltage:
Lamp Kit, UV, with power supply, 115 VAC, 60 Hz............................1 ....................... each....................20828-00
Lamp Kit, UV, with power supply, 230 VAC, 50 Hz............................1 ....................... each....................20828-02
Required Standards
Benzotriazole Standard Solution, 500-mg/L........................................0.6 mL ............. 100 mL...............21413-42
Program 95% Confidence Limits of Distribution
30 3–5 mg/L benzotriazole
730 3–5 mg/L tolyltriazole
Program Portion of Curve ∆Abs ∆Concentration
30 Entire range 0.010 1 mg/L benzotriazole
730 Entire range 0.010 1 mg/L tolyltriazole