DR/2400
Nitrate
Nitrate_UniCell_Dimeth_Eng_Ody.fm Page 1 of 4
• Adjust the pH of preserved samples to between pH 3–10 before testing.
• Clean the outside of the vial before placing it in the instrument. Wipe with a damp towel, followed by a dry one.
• Use the standard solution or standard addition method to verify results (see Accuracy Check).
• See Section 3.2.4 Adjusting the Standard Curve on page 40 for information on adjusting the calibration curve.
• Make sure that the temperature of the sample and the reagents is between 20 and 24 °C. Higher temperatures can cause
high-bias results; lower temperatures can cause low-bias results.
Nitrate
Dimethylphenol
UniCell™ Vials
Scope and Application: For water and wastewater
(1.0 to 60.0 mg/L NO
3
)
(0.2 to 13.5 mg/L NO
3
–N)
1. Touch
Hach Programs.
Select program
805 Nitrate, HCT 106.
Touch
Start.
2. Pipet 0.2 mL of
Dimethylphenol
Solution A (HCT 106 A)
into a sample vial.
3. Cap the vial and
invert 3 times to mix.
4. Immediately pipet
1.0 mL of sample into the
vial.
Tips and Techniques
UniCell Vials
Hach Programs

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach DR2400 and is the answer not in the manual?

Hach DR2400 Specifications

General IconGeneral
BrandHach
ModelDR2400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals