EasyManua.ls Logo

Hach DR2400 - Iron, Total

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DR/2400
Iron, Total
IronTot_AVPP_Other_FVR_Eng_Ody.fm Page 1 of 6
• Digestion is required for determining total iron for EPA reporting purposes. See the Hach Water Analysis Handbook for digestion
procedures.
• For more accurate results, determine a reagent blank value for each new lot of reagent. Follow the procedure using deionized
water in place of the sample. Subtract the reagent blank value from the final results or perform a reagent blank adjust
. See the
instrument manual for more information on Running a Reagent Blank.
• After adding reagent, an orange color will form if iron is present.
• Accuracy is not affected by undissolved powder.
Iron, Total
(Method 8008
FerroVer
®
Method*
*Adapted from Standard Methods for the Examination of Water and Wastewater
Powder Pillows or AccuVac
®
Ampuls
(0.02 to 3.00 mg/L)
Scope and Application: For water, wastewater, and seawater; digestion is required for determining
total iron; USEPA approved for reporting wastewater analysis**
** Federal Register, June 27, 1980; 45 (126:43459)
1. Touch
Hach Programs.
Select program
265 Iron, FerroVer.
Touch
Start.
2. Fill a clean, round
sample cell with 10 mL of
sample.
3. Add the contents of
one FerroVer Iron
Reagent Powder Pillow
to the sample cell (the
prepared sample). Swirl
to mix.
4. Touch the timer icon.
Touch
OK.
A three-minute reaction
period will begin.
(Allow samples that
contain rust to react for at
least 5 minutes.)
Tips and Techniques
Powder Pillows Method 8008
Hach Programs

Table of Contents

Related product manuals