EasyManua.ls Logo

Hach DR2400 - Cyanuric Acid

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DR/2400
Cyanuric Acid
CyanuricAcid_PP_Other_TUR_Eng_Ody.fm Page 1 of 4
• Filter highly turbid samples with filter paper (Cat. No. 1894-57) and a funnel (Cat. No.1083-67).
• After adding the reagent, a white turbidity will form if cyanuric acid is present.
• Clean sample cells with soap, water, and a brush soon after each test to avoid a build-up of film on the sample cell.
Cyanuric Acid
Method 8139
Turbidimetric Method
Powder Pillows
(5 to 50 mg/L)
Scope and Application: For water
1. Touch
Hach Programs.
Select program
170 Cyanuric Acid.
Touch
Start.
2. Fill a round sample
cell with 25 mL of
sample.
3. Add the contents of
one Cyanuric Acid 2
Reagent Powder Pillow
(the prepared sample).
Swirl to mix.
4. Touch the timer icon.
Touch
OK.
A three-minute reaction
period will begin.
Tips and Techniques
Powder Pillows Method 8139
Hach Programs
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL

Table of Contents