EasyManua.ls Logo

Hach DR2400 - Page 276

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cyanide
© Hach Company, 2003. All rights reserved. Printed in the U.S.A. 3/04 1ed
Page 4 of 4 Cyanide_UniCell_Other_PPZ_Eng_Ody.fm
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:
Call 800-227-4224
Contact the HACH office or distributor serving you.
www.hach.com techhelp@hach.com
In the U.S.A.
Outside the U.S.A.
On the Worldwide Web – ; E-mail –
toll-free
Method Performance
Precision
Standard: 0.30 mg/L Free CN
See Section 3.4.3 Precision on page 44 for more information, or if the standard
concentration did not fall within the specified range.
Sensitivity
See Section 3.4.5 Sensitivity on page 44 for more information.
Summary of Method
Cyanides react with chlorine to form cyanogen chloride, which in turn reacts
with pyridine in the presence of barbituric acid, condensing to form a violet-
colored compound. Measurements are taken at 588 nm.
Required Reagents
Description Unit Cat. No.
Cyanide - CN, UniCell™ HCT 129
*
................................................................................. 23/pkg.............. HCT 129
Required Apparatus
Adapter, 16-mm Cell .......................................................................................................... each....................59457-00
Optional Apparatus
Flask, volumetric, 100-mL ................................................................................................. each....................14574-42
Graduated cylinder, mixing, 100-mL............................................................................... each....................20886-42
Pipettor, (Jencons) 1–5 mL................................................................................................. each....................27951-00
Replacement tips for 27951-00 .......................................................................................... pk/100...............27952-00
Pipettor, (Jencons) 100–1000 µL........................................................................................ each....................27949-00
Replacement tips for 27949-00 .......................................................................................... pk/400...............27950-00
pH Paper.............................................................................................................................. pk/100...............26013-00
Optional Reagents
Potassium Cyanide............................................................................................................. 125 g.......................767-14
Sulfuric Acid Standard Solution, 1.00 N ......................................................................... 1 L ........................1270-53
Program 95% Confidence Limits of Distribution
813 0.23–0.37 mg/L CN
Program Portion of Curve Abs Concentration
813 Entire range 0.010 0.01 mg/L CN
* Available in Europe only

Table of Contents

Related product manuals