EasyManua.ls Logo

Hach DR2400 - Storing and Handling Reagents; Interpretation; Sample Collection and Storage

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alachlor
Alachlor
Page 6 of 8
ALA_None_Other_IMM_Eng_Ody.fm
Interpretation
Sample #1 — Sample reading is less than the readings for both calibrators.
Therefore the sample concentration of Alachlor is greater than both 0.1 ppb and
0.5 ppb Alachlor.
Sample #2 — Sample reading is between the readings for the 0.1 ppb and
0.5 ppb Alachlor calibrators. Therefore the sample concentration of Alachlor is
between 0.1 ppb and 0.5 ppb.
Sample #3 — Sample reading is greater than the readings for both calibrators.
Therefore the sample concentration of Alachlor is less than both 0.5 ppb and
0.1 ppb.
Storing and Handling Reagents
Wear protective gloves and eyewear.
When storing reagent sets for extended periods of time, keep them out of
direct sunlight. Store reagents at a temperature of 4 °C when not in use.
Keep the foil pouch containing the Antibody Cuvettes sealed when not in
use.
If Stop Solution comes in contact with eyes, wash thoroughly for 15 minutes
with cold water and seek immediate medical help.
Sensitivity
The Alachlor immunoassay test cannot differentiate between certain herbicides
and metabolites, but it detects their presence to differing degrees. The following
table shows the required concentration for selected chemicals.
Sample Collection and Storage
Collect samples in a clean glass bottle. Do not pre-rinse the bottle with the
sample. If the sample cannot be analyzed immediately, store the sample at 4 °C.
Samples may be kept for as long as 14 days. Warm the samples to room
temperature before analysis.
Compound
Concentration to give a
positive response of
0.1 ppb Alachlor
Concentration to give a
positive response of 0.5
ppb Alachlor
Acetochlor 0.45 ppb 4 ppb
Butachlor 0.09 ppm 1 ppm
2 Chloro-2',6'-Diethylacetaniline 0.030 ppm 2 ppm
Metolachlor 0.085 ppm 2 ppm
2,6-Diethylaniline 0.3 ppm 9 ppm
Propachlor 0.72 ppb 12 ppb
The following compounds are not detectable at 10,000 ppb:
Atrazine Carbofuran Carbendazim
Aldicarb 2, 4-D
Diazoton Chlorpyirfos

Table of Contents

Related product manuals