Boron
© Hach Company, 2003. All rights reserved. Printed in the U.S.A. 3/04 1ed
Page 4 of 4 Boron_PP_Other_CAR_Eng_Ody.fm
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:
Call 800-227-4224
Contact the HACH office or distributor serving you.
www.hach.com techhelp@hach.com
In the U.S.A. –
Outside the U.S.A. –
On the Worldwide Web – ; E-mail –
toll-free
Method Performance
Precision
Standard: 10.0 mg/L B
See Section 3.4.3 Precision on page 44 for more information, or if the standard
concentration did not fall within the specified range.
Sensitivity
See Section 3.4.5 Sensitivity on page 44 for more information.
Summary of Method
Boron is determined by its reaction with carminic acid in the presence of sulfuric
acid to produce a reddish to bluish color. The amount of color is directly
proportional to the boron concentration. Test results are measured at 605 nm.
Required Reagents
Quantity Required
Description Per Test Unit Cat. No.
BoroVer
®
3 Boron Reagent Powder Pillows .........................................1 pillow........... 100/pkg.............14170-99
Sulfuric Acid, ACS, concentrated...........................................................75 mL .............. 2.5 liters .................979-09
Water, deionized .......................................................................................2.0 mL ............. 4 liters....................272-56
Required Apparatus
Cylinder, graduated, 50-mL ....................................................................1 ....................... each........................508-41
Cylinder, graduated, 100-mL ..................................................................1 ....................... each........................508-42
Flask, Erlenmeyer, 125-mL ......................................................................2 ....................... each........................505-43
Flask, Erlenmeyer, 250-mL ......................................................................1 ....................... each........................505-46
Pipet, volumetric, 2.0-mL ........................................................................2 ....................... each....................14515-36
Sample cells, 10-20-25 mL, w/cap..........................................................2....................... 6/pkg.................24019-06
Required Standards
Boron Standard Solution, 10-mL Voluette
®
Ampule, 250-mg/L B ............................. 16/pkg...............14249-10
Program 95% Confidence Limits of Distribution
40 9.5–10.5 mg/L B
Portion of Curve ∆Abs ∆Concentration
0.010 mg/L 0.010 0.14 mg/L B
6 mg/L 0.010 0.15 mg/L B
13 mg/L 0.010 0.17 mg/L B