Chloramine (Mono)
Chloramine (Mono)
Page 2 of 6
Chloramine (Mono)_AVPP_High_MCF_Eng_Ody.fm
Interferences
The following have been tested for interference and found not to interfere up to
the indicated levels:
5. Remove the vial from
the adapter. Using a
micro-funnel, add the
contents of one
Monochlor F pillow to
the sample. Cap and
shake the cell about
20 seconds to dissolve.
6. Touch the timer icon.
Touch
OK.
A 5-minute reaction
period will begin.
7. After the color has
developed fully, place the
vial into the cell adapter.
8. Touch Read.
Results will appear in
mg/L Cl
2
.
Read
Table 1 Non-interfering Substances
Substance Maximum Level Tested
Alanine 1 mg/L N
Aluminum 10 mg/L
Bromide 100 mg/L Br
Bromine 15 mg/L Br
2
Calcium 1000 mg/L as CaCO
3
Chloride 18,000 mg/L
Chlorine Dioxide 5 mg/L ClO
2
Chromium (III) 5 mg/L
Copper 10 mg/L
Cyanide 10 mg/L CN
Dichloramine 10 mg/L as Cl
2
Fluoride 5 mg/L
Free Chlorine 10 mg/L Cl
2
Glycine 1 mg/L N
Iron (II) 10 mg/L
Iron (III) 10 mg/L
Magnesium 1000 mg/L as CaCO
3
Manganese (VII) 10 mg/L
Lead 10 mg/L

Table of Contents