Chloramine (Mono)
Chloramine (Mono)
Page 2 of 6
Chloramine (Mono)_MP_Low_MCF_Eng_2400.fm
Interferences
The following have been tested for interference and found not to interfere up to
the indicated levels:
5. Remove the cell from
theadapter and add the
contents of one pillow
Monochlor-F to the
sample. Cap and shake
the cell about 20 seconds
to dissolve.
6. Touch the timer icon.
Touch
OK.
A 5-minute reaction
period will begin.
Note: Samples colder than
18 °C will require additional
time. See Table 3 on page 3.
7. After the color has
developed fully, wipe the
cell and place it into the
adapter.
8. Touch Read.
Results will appear in
mg/L Cl
2
.
Read
Table 1
Substance Maximum Level Tested
Alanine 1 mg/L N
Aluminum 10 mg/L
Bromide 100 mg/L Br
Bromine 15 mg/L Br
2
Calcium 1000 mg/L as CaCO
3
Chloride 18,000 mg/L
Chlorine Dioxide 5 mg/L ClO
2
Chromium (III) 5 mg/L
Copper 10 mg/L
Cyanide 10 mg/L CN
Dichloramine 10 mg/L as Cl
2
Fluoride 5 mg/L
Free Chlorine 10 mg/L Cl
2
Glycine 1 mg/L N
Iron (II) 10 mg/L
Iron (III) 10 mg/L
Lead 10 mg/L
Nitrate 100 mg/L N
Nitrite 50 mg/L N