Metolachlor
Metolachlor
Page 2 of 8
MET_None_Other_IMM_Eng_Ody.fm
1. Touch
Single Wavelength
then touch the λ button.
Type in 450 nm and touch
OK.
2. Label an Antibody
Cuvette for each
calibrator and each
sample to be tested.
Note: As many as 20 cuvettes
may be tested at one time and
may comprise any combination
of samples and calibrators.
3. Place the cuvettes
into the rack snugly.
4. Pipet 0.5 mL of each
calibrator into the
appropriately labeled
cuvette.
Note: Use a new pipette tip for
each calibrator.
5. Pipet 0.5 mL of each
sample to be tested into
the appropriately labeled
cuvette.
Note: Use a new pipette tip for
each sample.
6. Immediately pipet
0.5 mL of Metolachlor
Enzyme Conjugate into
each cuvette.
7. Touch the timer icon.
Enter 20 minutes and
touch
OK.
A 20-minute reaction time
will begin.
Immediately mix the
contents of the cuvettes
for 30 seconds using the
technique described
in Using the 1-cm
MicroCuvette Rack on
page 5.
8. After 10 minutes mix
the contents of the rack
for 30 seconds using the
technique described
in Using the 1-cm
MicroCuvette Rack on
page 5.
Immunoassay Procedure for Water
λ
OK
54 0
HRS MIN SECHRS MIN SEC