EasyManua.ls Logo

Hach DR2400 - Page 394

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Molybdenum, Molybdate
Molybdenum, Molybdate
Page 2 of 6
Molybdenum_AVPP_High_MAA_Eng_Ody.fm
5. Add the contents of
one MolyVer 3 Reagent
Powder Pillow. Swirl to
mix (the prepared
sample).
6. Touch the timer icon.
Touch
OK.
A five-minute reaction
period will begin.
7. When the timer
beeps, fill a second cell
with 10 mL of the original
sample (the blank).
8. Insert the blank into
the cell holder.
9. Touch Zero.
The display will show:
0.0 mg/L Mo
6+
10. Place the prepared
sample into the cell
holder.
11. Touch Read.
Results will appear in
mg/L Mo
6+
.
10 mL
Zero
10 mL
Read

Table of Contents

Related product manuals