Nickel
Nickel
Nickel_PP_Other_HEP_Eng_Ody.fm Page 5 of 6
Method Performance
Precision
Standard: 1.00 mg/L Ni
See Section 3.4.3 Precision on page 44 for more information, or if the standard
concentration did not fall within the specified range.
Sensitivity
See Section 3.4.5 Sensitivity on page 44 for more information.
Summary of Method
Nickel ion reacts with heptoxime to form a yellow-colored complex which is
then extracted into chloroform to concentrate the color and enable a more
sensitive determination. Chelating agents are added to the sample to overcome
the interferences caused by cobalt, copper and iron. Readings are taken at
430 nm.
Required Reagents
Quantity Required
Description per test Unit Cat. No.
Nickel Reagent Set (50 Tests) .........................................................................................................................22435-00
Includes:
(3) Chloroform, ACS..........................................................................30 mL .............. 500 mL...............14458-49
(2) Nickel 1 Reagent Powder Pillows..............................................1 pillow........... 25/pkg.................2123-68
(2) Nickel 2 Reagent Powder Pillows..............................................1 pillow........... 25/pkg.................2124-68
Required Apparatus
Clippers, for opening powder pillows ..................................................1 ....................... each........................968-00
Cotton balls, absorbent ............................................................................1 ....................... 100/pkg...............2572-01
Cylinder, graduated, 10-mL ....................................................................1 ....................... each........................508-38
Cylinder, graduated, 500-mL ..................................................................1 ....................... each........................508-49
Funnel, separatory, 500-mL.....................................................................1 ....................... each........................520-49
Ring, support, 4-inch................................................................................1 ....................... each........................580-01
Sample Cells, 10-20-25 mL, w/cap.........................................................2....................... 6/pkg.................24019-06
Stand, support, 5" X 8" base ....................................................................1 ....................... each........................563-00
Stopper, hollow, Size 1 .............................................................................2 ....................... 6/pkg.................14480-00
Required Standards
Nickel Standard Solution, 1000-mg/L Ni....................................................................... 100 mL...............14176-42
Nickel Standard Solution, 300-mg/L Ni, 10-mL Voluette
®
Ampules ........................ 16/pkg...............14266-10
Water, deionized ................................................................................................................. 4 liters ....................272-56
Program 95% Confidence Limits of Distribution
335 0.93–1.07 mg/L Ni
Portion of Curve Abs Concentration
Entire range 0.010 0.02 mg/L Ni