EasyManua.ls Logo

Hach DR2400 - Page 494

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Nitrogen, Ammonium
© Hach Company, 2003. All rights reserved. Printed in the U.S.A. 3/04 1ed
Page 4 of 4 NitrogenAmm_UniCell_Low_SAL_Eng_Ody.fm
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:
Call 800-227-4224
Contact the HACH office or distributor serving you.
www.hach.com techhelp@hach.com
In the U.S.A.
Outside the U.S.A.
On the Worldwide Web – ; E-mail –
toll-free
Method Performance
Precision
Standard: 1.00 mg/L NH
4
+
–N
See Section 3.4.3 Precision on page 44 for more information, or if the standard
concentration did not fall within the specified range.
Sensitivity
See Section 3.4.5 Sensitivity on page 44 for more information.
Summary of Method
Ammonium ions react with hypochlorite ions and salicylate ions at pH 12.6 in
the presence of sodium nitroferricyanide to form indophenol blue. Test results
are measured at 694 nm.
Required Reagents
Description Unit Cat. No.
Ammonium - NH
4
,
UniCell™ HCT 100.......................................................................... 23/pkg.............. HCT 100
Optional Reagents
Ammonium Standard Solution, 100-mg/L as NH
3
–N ................................................. 500 mL...............24065-49
Nitrogen Ammonia Standard, 1-mg/L as NH
3
–N........................................................ 500 mL.................1891-49
Sodium Hydroxide Standard Solution, 5.0 N................................................................. 50 mL SCDB .......2450-26
Sulfuric Acid, ACS.............................................................................................................. 500 mL...................979-49
Required Apparatus
Adapter, 16-mm Cell .......................................................................................................... each....................59457-00
Optional Apparatus
Cylinder, mixing, 100-mL.................................................................................................. each....................20886-42
Flask, volumetric, 1000-mL ............................................................................................... each....................14574-53
Pipettor, Jencons, 1–5 mL................................................................................................... each....................27951-00
Replacement tips for 27951-00 .......................................................................................... 100/pkg.............27952-00
Program 95% Confidence Limits of Distribution
802 0.76–0.80 mg/L NH
4
+
–N
Program Portion of Curve Abs Concentration
802 Entire range 0.010 0.01 mg/L NH
4
+
–N

Table of Contents

Related product manuals