EasyManua.ls Logo

Hach DR2400 - Page 739

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DR/2400
Sulfate
Sulfate_UniCell_Other_Turb_Eng_Ody.fm Page 1 of 4
• Adjust the pH of preserved samples to between pH 2–10 before testing.
• Clean the outside of the vial before placing it in the instrument. Wipe with a damp towel, followed by a dry one.
• Use the standard solution or standard addition method to verify results (See “Accuracy Check” on page 2.).
• See Section 3.2.4 in the procedure manual for information on adjusting the calibration curve.
• Make sure that the temperature of the water sample and the sample vial is between 15–25 °C (59–77 °F).Temperatures outside of
this range may give inaccurate results.
Underrange appears on the display when the instrument is zeroed or when the determined sample concentration is below the
operating range listed for this method.
Sulfate
Turbidimetric Method
UniCell™ Vials
Scope and Application: For wastewater, raw water, and process control
(40 to 150 mg/L SO
4
2–
)
1. Touch
Hach Programs.
Select program
811 Sulfate, HCT 125.
Touch
Start.
2. Add 5 mL of sample
to a sample vial.
3. Add one spoonful
of Barium Chloride A
(HCT 125 A) into the
sample vial.
4. Immediately cap the
sample vial and invert to
mix. Mix for 2 minutes.
Tips and Techniques
UniCell Vials
Hach Programs

Table of Contents

Related product manuals