Chlorine, Free and Total, HR, continued
4.Press: ZERO/SCROLL
The display will show
“- - - -” followed by “0.0”.
Remove the blank.
5.Fill another 1-cm/10-mL
sample cell to the 5-mL line
with sample. Cap.
Note: Do not use the same
sample cells for free and
total chlorine without
thoroughly rinsing the cells
between the free and total
tests.
6.Use the SwifTest™
Dispenser to add two
dispensations of DPD Free
Chlorine reagent or DPD
Total Chlorine reagent to
the sample cell
(the prepared sample).
Cap the cell and shake
gently for 20 seconds.
Note: Gentle shaking
dissipates bubbles.