2—11
Error Codes, continued
Verify that the concentration of the standard has been entered correctly.
If the problem persists, contact a Service Center (page 2—37).
5. E-6 Abs Error (User mode)
Indicates that the absorbance value is invalid, or indicates an attempt to
make a curve with less than two points.
Enter or measure the absorbance value again.
If the problem persists, contact a Service Center (page 2—37).
6. E-7 Standard Value Error (User mode)
Standard concentration is equal to another standard concentration that is
already entered.
Enter the correct standard concentration.
If the problem persists, contact a Service Center (page 2—37).