2—12
Error Codes, continued
7. E-9 Flash Error
The meter is unable to save data.
If the problem persists, contact a Service Center (page 2—37).
8. Underrange—flashing number below stated test range
Verify instrument cap is correctly seated.
Check zero by measuring a blank. If error recurs, re-zero the
instrument.
If the problem persists, contact a Service Center (page 2—37).
9. Overrange—flashing number above stated test range
Note: Flashing value will be 10% over the upper test limit.
Check for light blockage.
Dilute and retest sample.