Maintenance and Diagnostics
9.8.2 Priming tubes
9.8.3 Bottles full
After refilling the bottles (reference electrolyte, calibration
solution and reactivation solution) and/or after exchanging the
empty bottle of conditioning reagent with a new one, select
the PRIMING TUBES option.
This will set off the process to purge and refill the tubes.
A progress bar is displayed while the process executes.
Press the Esc function key at any time to abort the process
and return to the menu.
If the bottles of reagent are full, select Yes and the default
values for the volumes will be set. These values can be
viewed and modified by selecting the ADJUST BOTTLE
VOLS. option in the menu.
If you select No to this option, you will manually have to enter
the correct values using the ADJUST BOTTLE VOLS. option.