14
Specifications
MECHANICAL SPECIFICATIONS
Maximum panel dimension
(H x L x D)
850 x 450 x 252.5mm [33.46 x 17.71 x 9.94in]
Inlet
Simple fittings for 6 mm O.D. tubing or ¼" O.D. in
PE-low density. ¼" OD in PHED-PTFE-SS as option
Outlet Barbed stem for 12 mm (½" I.D.) hose
Protection rate
Transmitter: IP65 (NEMA 4)
Panel: IP50 (Dust protection)
Optional Enclosure: IP54 (Splash water proof),
Instrument is designed to avoid DIPA vapor inside the enclosure.
All DIPA vapor is collected and sent to the instrument drain
Cell PMMA - compact (minimum tubing)
Flame rate Conform UL
Maximum weight 15 - 30 Kg
INTERFACE SPECIFICATIONS
Mains power supply 100 - 240 VAC, 50-60 Hz, ± 10%, automatic switching
Max. consumption 80 VA.
Fuse 5 x 20 cartridge T2AL-250V following CEI127
Display
Curve trend, Last Cal Date, Historical, Concentration,
Temperature, Potential
Analog outputs
Number: 6
4-20 or 0-20 mA (650 ohms)
Linear / Dual / Logarithm
Event indication
Relays
4 x Relay (conc)
1 x Warning
1 x System
Logic input
Active / Inactive channels
Remote AutoCal
STANDARDS
EMC requirements
EMC Directive 2004/108/EC; EN 61326-1
NOTICE
This is a Class A product intended for use in an industrial
environment. In a domestic environment this product may cause
radio interference in which case the user may be required to take
adequate measures.
NOTICE
Measurement variations of less than 5% of the full range can
occur if the instrument is subject to a strong electromagnetic field.
European safety standards EN 61010-1 for low voltage safety
International standards cETLus
1
Experiment performed between ultra pure water (estimated at 50 ppt) and 1 ppb