7
Specifications
Table 2 Individual electrical specifications
Module
AC input voltage
rating or range
AC input frequency
rating or range
Maximum active
power (W),
apparent power
(VA) or current
(A)
rating
Fuse rating
PDS200 110–240 VAC 50–60 Hz 0.5 A T, ½ A, 250 V
ASX-500 sampler 100–240 VAC 50–60 Hz 1.5 A
Fuse not
user-replaceable
ASX-400 sampler 100–120 VAC 50–60 Hz 200 VA
Fuse not
user-replaceable
Reagent pump
100 or 120 VAC 50–60 Hz 15 VA T, ½ A, 250 V
220 or 240 VAC 50–60 Hz 15 VA T, ¼ A, 250 V
System unit
100–115 V ± 10% 50–60 Hz 800 W T, 8 A, 250 V
230 V ± 10% 50–60 Hz 800 W T, 4 A, 250 V
Table 3 Environmental specifications
Temperature 15–35 °C (60–95 °F)
Use Laboratory
Altitude Up to 10,000 feet (3 km) above sea level
Relative humidity Up to 92.5% non-condensing
Pollution degree II
Installation category II
Table 4 Certifications
Certification Marks Standards
CE
IEC/EN 61010-1 per LV Directive (2006/95/EC), IEC/EN 61326 per EMC: Directive
(2004/108/EC
cETL CSA C22.2 No 61010-1
ETLus UL 61010-1