EasyManua.ls Logo

Hach Sigma 900 MAX - Page 49

Hach Sigma 900 MAX
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
49
Basic programming setup
1-C. Enter the bottle volume and select gallons or milliliters using the CHANGE UNITS
key.
1-D. Press
ACCEPT and continue to Intake Tubing.
Step 2 - Intake tubing
2-A. Enter the intake tube length of the intake tubing attached to the sampler. Length
values from 100 to 3000 cm (3 to 99 ft) are valid. Change the measurement unit
using the
CHANGE UNITS key.
Note: Intake tube length affects sample volume accuracy, cut the tubing to the nearest whole foot.
Avoid excessively long tubing runs. Doubling the intake tubing length can quadruple the pump tubing
wear if intake rinses or sample retries are enabled. Always locate the sampler close to the sample
liquid source to minimize tubing, pump, motor, and gearbox maintenance.
2-B. Press ACCEPT to move to the Intake Tube Type menu.
2-C. Select the type of intake tube (
3
/8 in. Vinyl, ¼ in. Vinyl,
3
/8 in. Teflon
®
) using CHANGE
UNITS
key.
2-D. Press
ACCEPT to select the intake tube type and continue with Program Lock.
Step 3 - Program lock
Note: The program lock password is configured at the factory as “9000” and cannot be changed.
Enabling the Program Lock will provide a protective “password” which will keep
unauthorized personnel from tampering with the instrument keypad.
3-A. Enable or Disable Program Lock using the
CHANGE CHOICE key.
Note: When program lock is enabled and a user attempts to make a change in the program, a
screen will ask the operator to enter the password. The operator must enter 9000 and press
ACCEPT.
3-B. Press ACCEPT to continue the basic program setup and continue with Program
Delay.
11:00 AM 21 - APR - 01 BOTTLES
ACCEPT
BOTTLE VOLUME
CHANGE
UNITS
— —
BACKUP
CLEAR
ENTRY
ENTER: 0.500—99.90
11:00 AM 21 - APR - 01 INTAKE TUBING
ACCEPT
INTAKE TUBE LENGTH:
CHANGE
UNITS
_____
BACKUP
CLEAR
ENTRY
ENTER: 3—99
Goodnal STP ST041 Pre Treatment (Sigma 900 MAX All Weather Refrigerated Sampler) Vendor Manual
Q-Pulse Id VM373
Active 29/10/2013
Page 49 of 122

Table of Contents

Related product manuals