SPECIFICATIONS, continued
11
OPTICAL SPECIFICATIONS
Light Source: Class 1 LED (Light Emitting Diode) with peak wavelength of
610 nm. Estimated 50,000 hours minimum life.
Detector: Silicon photo detector
Optical Path Length: 0.35"
MECHANICAL SPECIFICATIONS
Enclosure: Environmental Rating IP62
Dimensions: Approximately 42 x 31.5 x 18 cm (16.5 x 12.5 x 7") (refer to the
transmittal drawing for exact space requirements).
Case Construction: IP62 instrument enclosure with gasketed door
Mounting: Wall-mounted
Shipping Weight: 11.3 kg (25 lbs)
HYDRAULIC (SAMPLING) SPECIFICATIONS
Sample Flow Rate to Sample Conditioning: 50 to 500 mL/minute flow
rate required.
Inlet Pressure to Instrument: 1 to 5 psig, 1.5 psig is optimum. Exceeding 5 psig
can cause sample tubing failure.
Inlet Pressure to Sample Conditioning: 1.5psigto75psig(withthesample
tube level with the bottom of the instrument)
Inlet Fitting:
At instrument, ¼" OD polyethylene tubing with quick-disconnect fitting
Drain Fitting: Hose barb for ½" ID flexible tubing
Sample Temperature Range: Cto4C(41°Fto10F)
Sample Conditioning: use provided sample conditioning