8-8 Consulting Reference Information
P/N 85213-30, Manual revision: B
IEC 60601-1-2, Table 202: Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment guidance
*Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least
30%.
*Electrical fast
transient burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 2 kV for power
supply lines
Mains power quality should be that of a typi-
cal commercial or hospital environment.
*Surge
IEC 61000-4-5
± 1 kV differen-
tial mode
± 2 kV common
mode
± 1 kV differen-
tial mode
± 2 kV common
mode
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Voltage dips,
short interrup-
tions, voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% U
T
(> 95%
dip in U
T
) for 0.5
cycles
40% U
T
(60%
dip in U
T
)
for 5 cycles
70% U
T
(30%
dip in U
T
)
for 25 cycles
< 5% U
T
(> 95%
dip in U
T
) for 5
sec
< 5% U
T
(> 95%
dip in U
T
) for 0.5
cycles
40% U
T
(60%
dip in U
T
)
for 5 cycles
70% U
T
(30%
dip in U
T
)
for 25 cycles
< 5% U
T
(> 95%
dip in U
T
) for 5
sec
Mains power quality should be that of a typi-
cal commercial or hospital environment.
If the operator of the Haemonetics system
requires continued operation during power
mains interruptions, it is recommended that
the Haemonetics system be powered from an
uninterruptible power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.
* Authorized maximum value

Table of Contents

Other manuals for Haemonetics MCS+

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Haemonetics MCS+ and is the answer not in the manual?

Haemonetics MCS+ Specifications

General IconGeneral
BrandHaemonetics
ModelMCS+
CategoryMedical Equipment
LanguageEnglish

Related product manuals