Haemonetics recommends the following operator actions to remove any air
bubbles detected in the tubing between the BLAD, DLAD1 and the DLAD2.
The operator should:
" Press the Draw key until blood enters the bowl to send any air bubbles to
the bowl.
" Continue with a RETURN cycle only after any air bubbles have been re-
moved.
Warning: If, after attempting this procedure to remove any air, the DLAD1 and/
or DLAD2 detect air again, the operator should terminate the procedure, dis-
continue use of the device and contact the authorized Haemonetics represen-
tative.
Disposable set
element holders
Located on the MCS+ front panel are the following various components used to
secure elements from the disposable set during a collection procedure:
! Blood filter chamber brackets.
! Re-circulation chamber brackets.
! Disposable set pins.