8-10 Consulting Reference Information
P/N 85213-30, Manual revision: B
The Haemonetics system is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the
Haemonetics system can help prevent electromagnetic interference by maintain-
ing a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Haemonetics system as recommended below,
according to the maximum output power of the communications equipment.
IEC 60601-1-2, Table 206: Recommended separation distance between portable RF communications
equipment and the Haemonetics device - Non-live supporting equipment
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.21.22.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be esti-
mated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz, the separation for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
d 1.2 P= d 1.2 P=
d 2.3 P=