90
FLUORIDE HIGH RANGE
Note: For wastewater or seawater samples, before performing measurements, distillation is
required.
INTERFERENCES
Interferences may be caused by:
Alkalinity (as CaCO
3
) above 5000 mg/L
Aluminum above 0.1 mg/L
Iron, ferric above 10 mg/L
Chloride above 700 mg/L
ortho‑Phosphate above 16 mg/L
Sodium hexametaphosphate above 1.0 mg/L
Sulfate above 200 mg/L
Highly colored and turbid samples may require distillation
Highly alkaline samples can be neutralized with nitric acid.