7.2. pH SETUP
To access pH Setup, press option key while in pH mode.
Use and keys to highlight the desired option.
Press or to access the selected option.
7.2.1. BUFFER ENTRY TYPE
Option: Automatic, Semiautomatic or Manual
Automatic: The instrument automatically selects the pH calibration point as the closest buffer from the predefined buffer group (see
Edit Buffer Group section).
Semiautomatic: The instrument automatically selects the closest buffer from the available buffers (standard and custom buffers).
Manual: The calibration buffer must be manually selected by the user during calibration from the available buffer list (standard and
custom buffers).