ISE MODE
Press from the main screen. If the instrument was calibrated before and the calibration was not cleared, the old
calibration can be cleared by pressing .
Immerse the ISE and the temperature probe approximately 2 cm into the standard with the lowest concentration.
Select the standard concentration with or .
When the reading has stabilized, press to update the calibration. The calibration point value will be added to the
Calibrated Standard list.
• Select and repeat the procedure with all of the available standards.
• Press to exit the calibration.
9.4. LOGGING
Data logging is available in ISE mode. It can be logging on demand (Manual Logging) or automatically (Interval Logging) at
predefined time intervals.
To customize the logging report:
• Press to display the Data Parameters screen.
Highlight the Setup pH/mV/ISE Report option and press to display the Setup pH/mV/ISE Report screen.
Use the and keys to highlight the data field that you want to show/hide in the pH/mV/ISE report and then press
to activate/deactivate it.
Each field marked by “*” is an active field selected for the report.
• Press to save the customized report.