10-3
AUXILIARY FUNCTIONS
10.1. BURETTE
To access the Burette screen, press from the main titration screen.
Highlight the desired option and then press .
allows you to select the desired pump for burette operations (it is only active if two pumps are connected).
10.1.1. PRIME BURETTE
Option: Up to 5
The Prime Burette option is used to fill the burette with titrant or reagent before starting a titration. The priming process consists of
several cycles of filling and emptying the burette with titrant.
Two rinse cycles of burette are shown in the figure below. The dispensing tube is connected on the right side and the aspiration tube
on the left side.