10-4
AUXILIARY FUNCTIONS
Note: Before starting this operation, the aspiration tube must be inserted in the titrant bottle. A waste container should be placed
under the dispensing tip to collect the waste solution.
To prime the burette, select Prime Burette, enter the number of rinses and press .
We recommend at least three rinses to assure that the air bubbles are completely removed.
10.1.2. RINSE TIP
A 2 mL dose of titrant will be dispensed from the burette when this operation is selected, this will eliminate any air in the dispensing tip.
10.1.3. MANUAL DISPENSE
Manual Dispense option allows a defined titrant volume to be dosed. Select the Manual Dispense option and press .