4-12
GENERAL OPTIONS
4.13. SETUP BALANCE INTERFACE
This screen allows the users to connect an analytical balance for automatic acquisition of sample mass prior to titration or
standardization.
The balance is connected to the titrator via RS 232 interface.
Press to add a new balance to the list.
Press to enable the selected balance.
Press to disable the selected balance (automatic weight acquisition will be not available).
Press to customize the name and serial communication parameters used by the selected balance.
Be sure that the settings on the titrator Balance Configuration menu match the settings for your particular balance (baud rate, data
bits, parity, stop bits number, request command syntax). It may be necessary to change settings on your balance. Users should
consult their balance instruction manual.
Before leaving this screen, be sure the connection with the balance is working properly by pressing the key.