5-12
TITRATION METHODS
Fixed End Point (mV):
Option: -2000.0 to 2000.0 mV
The end point detection algorithm is the same as for pH, but the threshold value is expressed in mV.
5.5.6.2. EQUIVALENCE END POINT (pH OR mV)
The titration is terminated when the equivalence point is detected (the point where the added quantity of titrant equals the quantity
of analyte present in the sample).
Number of Equivalence Points:
Option: 1 to 5