5.5.7. RECOGNITION OPTIONS (EQUIVALENCE END POINT ONLY)
The Recognition Options screen is a set of parameters used to avoid false detection of the equivalence point due to the chemical
system (titrant/sample species and concentrations) and/or electrode response.
5.5.7.1. THRESHOLD
Option: 1 to 9999 mV/mL
This parameter must be set by the user according to the analysis.
The threshold represents the absolute value of the first derivative, expressed in mV/mL, below which the detection algorithm does not
search for the equivalence point.
The recommended value is 40% of the absolute value of the first derivative.