5-24
TITRATION METHODS
A formula example is shown below using M (mol/L) as the titrant unit and ppt (g/kg) as the final result unit:
Variables can be set according to the amount of sample and titrant used.
5.5.13.1.5. SAMPLE CALCULATIONS BY VOLUME
This calculation is used when the concentration of an analyte is determined in terms of the volume of sample. The results are based
on the initial sample volume (in milliliters).
The titrator will calculate the results based on the selected units.
Titrant Units:
M (mol/L) moles/liter
N (eq/L) equivalents/liter
g/L grams/liter
mg/L milligrams/liter

Table of Contents