6-5
TITRATION MODE
6.2. STOPPING A TITRATION
The titration is terminated when one of the following conditions is met:
Titration Completed: This is the only mode with valid final result values. The end point/stable reading was successfully
detected, the final results will be displayed.
Manually Terminated: The current titration terminated by the user before the end point was detected.
Limits Exceeded: The maximum titrant volume was delivered without reaching the end point. An error message is displayed on
the screen.
Critical Error: A critical error occurred and the titration was stopped. These errors are typically related to the dosing system. An
error message is displayed on the screen.
Potential Out of Range: The measured values from the electrode are outside the potential range. An error message is
displayed on the screen.