39
7.3. FAVORITE METHODS
Frequently used methods can be tagged as a favorite method, see page 17 for additional information. Favorite methods can be both
factory and user methods. Up to 30 methods can be tagged as favorites.
Once a method has been tagged as a favorite it will appear in the Favorite Method list for easy access when the key is pressed.
Favorite Methods can also be easily accessed from the main screen by pressing the key.
7.4. CREATE NEW
This options allows users to develop new user methods.
Press the CFM key to create a new method. A series of prompts will guide the user through the process of creating a new method.
Press NEXT to proceed to the next setting.
Press BACK to return to the previous setting.
7.4.1. METHOD NAME
Option: Up to 12 alphanumeric characters
Use the or key to select the desired character. Press the or
key to move across characters.
Press the NEXT key to save and continue or the BACK key to return to the methods
menu.
For more information on the settings and options that are available during method creation, see section 6.4 Method Settings.
METHODS
After all settings have been entered, press the CFM key to create the method. The
meter will show "Method Created" before returning to the main screen.
All of these settings can be modified in method settings, see page 28 for additional
information.
In order to use the newly created method that reports in a concentration unit a
calibration must be done.
A calibration is not required for methods reporting in absorbance, % transmittance or
multiwavelength.