PREVENTATIVE MAINTENANCE GUIDE
FOR ELECTRIC HOSPITAL CRIBS
4.0 Performance Testing
YES NO
4.1 Is the green light on the under crib control box lit showing it is getting power?
4.2 Press the head up/down buttons allowing the head to travel to the top of its range
prior to pressing the down button. Is it functioning correctly and getting full range of
motion?
4.3 Press the foot up/down buttons allowing the foot/knee section to travel to the top
of its range prior to pressing the down button. Is it functioning correctly and getting full
range of motion?
4.4 Press the Hi-Lo up/down buttons allowing the mattress platform to travel to the top
of its range prior to pressing the down button. Is it functioning correctly and getting full
range of motion?
4.5 Does the scale power up and display the current weight (which may be zero)?
4.6 When a certi ed 10kg weight is placed on the crib, does the scale appear to
weigh correctly? Note- this is the only unit of measure that can be used to calibrate or
test a scale.
Sta Control, Transformer, Scale and Motors
Plug the crib into an electrical outlet if unplugged.
Scale Control if present
____ ____
____ ____
____ ____
____ ____
____ ____
____ ____