7
Pump body: the heart of the whole system and the driving force of transfusing medical liquid. It uses
step motor driving screw to push the syringe plunger forward.
Detection device: mainly containing sensors,such as ultrasonic sensor (for detecting motor running
and reversing) and pressure sensor (for detecting occlusion) etc. They can detect corresponding signals,
which after being amplified and transferred to microcomputer system for signal processing and thus incur
control instruction for corresponding operation.
Alarm system: The signals detected by the sensor, after being processed by the microcomputer, shall
incur alarm control signal and then at the response of alarm system, which alert the user for immediate
correct operation. It contains mainly photoelectric alarm (light emitting diode) and audible alarm
(loudspeaker and buzzer) etc.
HMI: to set parameters of infusion through keys, such as infusion volume and speed. It can display
all parameters and current working state through color LCD.
4. Technical Characteristics, Parameters and Explanations of Nouns
Technical and parameters
Applicable syringe
5, 10, 20, 30, 50/60ml disposable sterile syringes
Infusion mode
Rate mode
Time mode
Weight mode
Intermittent mode
Program Mode
TPN Mode
TIVA mode
Shift Mode
Volume to be Infused
(VTBI)
(0-9999.99)ml