8. Operation Method
8.1 Operation
8.1.1 Fix the machine and connect the power cord
Install the syringe pump and connect it to AC power. AC/DC indicator light will be on. The battery will
start charging when connect to external power.
8.1.2 Switch on/off
Switch on: Press Power key few seconds, the pump will be switched on and do self-test, and
shows Self-testing on the display, it will test: Communication Info, Press sensor, Potentiometer, Dedicated
IV set, AC Info and Battery Info. The test result (√ or ×) will be displayed on LCD.
Pump self-test information after switching on:
(1) Communication information: √ indicates CPU normal; × indicates CPU abnormal;
(2) Battery information: √ indicates internal battery normal; × indicates internal battery abnormal
(3) AC information: √ indicates AC connected well; × indicates AC fail;
(4) Pressure sensor: √ indicates normal pressure detection; × indicates: abnormal pressure
detection;
(5) Potentiometer: √ indicates normal Potentiometer; × indicates: abnormal Potentiometer;
(6) Dedicated IV set: √ indicates normal dedicated IV sensor; × indicates: abnormal installation of
dedicated IV set or IV set is installed in the pump already;
Attention: 1. If the potentiometer is abnormal, the system will ask “Is the syringe installed?”,
please choose the answer according to the facts, the system will predicate per your choice.
2. Please keep an eye on the self-test, if any abnormal, please contact our company, DO NOT
force to use it.
Press POWER key for about 2 seconds to turn off the machine. After the machine turn off,
the accumulated volume will become 0ml.
8.1.3 Install the syringe